SPECIFICATIONS
1 TOB-CT-4008-5V5mA Button Battery Tester is an eight channel battery analyzer to analyze small coin cells with a range current 0.025-5mah, voltage 0-5v. Each channel of the analyzer has independent constant-current and constant-voltage source, which can be programmed and controlled by computer software.
2 This Button Battery Tester provides most applications in battery testing fields such as electrode materials research, battery performance test, small scale battery formation, capability grading, battery pack testing, etc.
3 The Button Battery Tester is suitable for all kinds of coin cell battery such as lithium battery, NI-MH battery, Zn-Mn watch battery and so on.
4 The 4 system tester is with more quicker testing speed and add DRIR testing.
Project index |
Parameter index |
|
Input Power |
AC 110V/220V ±10% 50Hz/60Hz |
|
resolution |
AD: 16 bit; DA: 16 bit |
|
Input Resistance |
≥10 GΩ |
|
Input Power |
354 mW |
|
Channel Features |
Constant Current Source and Constant Voltage Source with Independent pairs of Closed-loop Structure |
|
Channel Control Mode |
Independent Control |
|
Voltage |
Constant Voltage Range |
25 mV~5 V |
Discharge Min Voltage |
0 V |
|
Accuracy |
± 0.05% of FS |
|
Stability |
± 0.05% of FS |
|
Current |
Per Channel Current Range |
0.025mA-5mA |
Accuracy |
± 0.05% of FS |
|
Stability |
± 0.05% of FS |
|
Power |
Per Channel Output Power |
0.025 W |
Stability |
± 0.1% of FS |
|
Time |
Current Response Time |
1ms (0-Full Range) |
Step Time Range |
≤ (365*24 ) hour/step, Time Form Support 00:00:00(h.min:s.ms) |
|
Data Recording |
Record Condition |
Minimum Time Interval: 100 ms |
Minimum Voltage Interval: 10 mV |
||
Minimum Current Interval: 0.01 mA |
||
Record Frequency |
10 Hz |
|
Charge |
Charge Mode |
Constant Current Charge, Constant Voltage Charge, Constant Current and Constant Voltage Charge, CPC |
End Conditon |
Voltage, Current, Relative Time, Capacity |
|
Discharge |
Discharge Mode |
CCD, CPD, CRD |
End Condition |
Voltage, Current, Relative Time, Capacity |
|
Pulse |
charge |
CCC, CPC |
Discharge |
CCD, CPC |
|
Min Pulse Width |
500 ms |
|
Automated Switch |
Automated swith from charge to discharge for each pulse |
|
End Condition |
Voltage, Test Time |
|
Cycle |
Loop Measure Range |
1~65535 times |
Max Steps Per Loop |
254 |
|
Nested Loop |
Nested loop Function, Max Support 3 Layers |
|
Protection |
Software Protection |
Power-down Data Protection |
Off-line Testing Function |
||
Setting Protection Condition,Setting Parameter: Low Voltage limit, Upper Voltage limit, Low Current limit, Upper Current limit, Delay Time |
||
Hardware Protection |
Additional Anti-reverse Protection Model |
|
DCIR testing |
Support |
|
Voltage and Current Testing Sample |
4-wire connecting |
|
Noise |
< 85 dB |
|
Database |
MySQL Database |
|
A Means of Communication of Upper Machine |
TCP/IP Protocol |
|
Data Output |
EXCEL, TXT, Graph |
|
Communication Interface |
Ethernet Port |
|
Per Unit Main Channels No. |
8 |
|
Dimensions |
250*200*60 (mm) |
Product Display