+7(495)722-60-73 [email protected] 117292, г.Москва, ул.Кедрова, д.4, корп.1
Conductivity Meter With Temperature Compensation -5-135℃

Conductivity Meter With Temperature Compensation -5-135℃

SPECIFICATIONS


This TOB-DDSJ-318 electrical conductivity meter Can test Electrical conductivity, Electrical resistivity, TDS, Salinity, Temperature for (-5.0~135.0) ℃

Model

Conductivity Meter With Temperature Compensation TOB-DDSJ-318

Source voltage

AC 220V/110V, 50/60HZ

Warranty

One year limited warranty with lifetime support

Advantages

1 Can test Electrical conductivity, Electrical resistivity, TDS, Salinity, Temperature

2 Adopt microprocessor technology, with automatic temperature compensation, automatic calibration, automatic switch range, etc

3 Big screen, blue backlight, double digital display LCD

4 With standard RS - 232 and USB interface, can connect with PC

Testing range

Electrical conductivity: 0.000 μS/cm~199.9 mS/cm

Electrical resistivity: 5.00 Ω.cm~20 MΩ.cm

TDS: 0.000 mg/L~19.99 g/L

Salinity: (0.0~8.00) %

Temperature:(-5.0~135.0)℃

Resolution ratio

Electrical conductivity: 0.001 μS/cm、0.1 mS/cm

Electrical resistivity: 0.1Ω.cm

TDS: 0.001 mg/L

Salinity: 0.01%

Temperature: 0.1℃

Testing error

Electrical conductivity: ±0.5% FS

Electrical resistivity: ±0.5% FS

TDS: ±0.5%FS

Salinity: ±0.1%

Temperature: ±0.3℃

Constant conductance pool with best conductivity measurement range

Electrode constant

(cm-1)

0.01

0.1

1

10

Test range

(μS/cm)

0.-19.99 μS/cm

2-199.9 μS/cm

20.0 μS/cm-19.9 ms/cm

200 μS/cm-199.9 ms/cm

Temperature compensation scope

(-5.0~135.0) ℃

Dimension (L*W*H)

250×195×100 mm

Weight

About 2 kg


footer_logo